Model | | MCR 72 | MCR 92 |
Specifications | Units | | |
Bearing | - | Ball | Air |
EC motor (brushless DC) | - | ✔ | ✔ |
with high resolution optical | | | |
encoder | | | |
Rotation Mode | - | ✔ | ✔ |
Oscillation Mode | - | ✔(1 | ✔ |
Direct Strain Controller | - | ✔ | ✔ |
Direct Stress Controller | - | ✔ | ✔ |
Maximum torque | mNm | 125 | 125 |
Minimum torque, rotation | µNm | 200 | 1 |
Minimum torque, oscillation | µNm | 200 | 1 |
Torque resolution | nNm | 100 | 100 |
Angular deflection, set value | µrad | 1 to ∞ | 1 to ∞ |
Angular deflection, resolution | nrad | 614 | 614 |
Step rate, time constant | ms | 100 | 100 |
Step strain, time constant | ms | 100 | 100 |
Minimum angular velocity (2 | rad/s | 10-4 | 10-4 |
Maximum angular velocity | rad/s | 157 | 157 |
Minimum angular frequency (3 | rad/s | 10-3 | 10-4 |
Maximum angular frequency | rad/s | 628 | 628 |
Minimum speed (CSS/CSR) | rpm | 10-3 | 10-3 |
Maximum speed | rpm | 1500 | 1500 |
Maximum temperature range | °C | -40 to 400 | -40 to 400 |
SafeGap, Normal force Limiter | - | ✔ | ✔ |
during Gapsetting | | | |
TruRay, Dimmable illumination | - | ✔ | ✔ |
of sample area | | | |
Connections | | USB, Ethernet, RS232, Analog interfaces, |
| | Pt100 port |
Dimensions | mm | 380 x 660 x 530 | 380 x 660 x 530 |
Weight | kg | 33 | 33 |
QuickConnect for measuring | - | ✔ | ✔ |
systems, screwless | | | |
ToolmasterTM, | - | ✔ | ✔ |
measuring system | | | |
ToolmasterTM, | - | ✔ | ✔ |
measuring cell | | | |
CoolPeltier, Peltier-controlled | °C | 25 below ambient temperature but not lower than -5 up to +200 (4 |
Plate System | | | |
with built in cooling option | | | |
that requires no additional | | | |
accessories for counter-cooling | | | |
Actively Peltier-controlled hood | °C | -5 to 200 (4 |
that requires no additional | | | |
accessories for counter-cooling | | | |
CoolPeltier, Peltier-controlled | °C | 15 below ambient temperature but not lower than +5 up to +150 (4 |
Cylinder System with built in | | | |
cooling option that requires | | | |
no additional accessories | | | |
for counter-cooling | | | |
Virtually gradient-free | | ✔ | ✔ |
(horizontal, vertical) | | | |
temperature control | | | |
Rheometer Software: | | | |
Test Designer | - | ✔ | ✔ |
Report Designer | - | ✔ | ✔ |
User Management | - | ✔ | ✔ |
Electronic trim lock for the | - | ✔ | ✔ |
measuring system | | | |
Automatic gap control/setting, | - | ✔ | ✔ |
AGC/AGS | | | |
| | | |
1) Depending on sample properties | | |
2) Depending on measuring point duration and sampling time, practically any value is achieved |
3) Set frequencies below 10-4 rad/s are of no practical relevance due to the measuring point |
duration > 1 day | | | |
4) System temperature, sample temperature may vary. For measurements at very high |
or low temperatures a calibration in the sample gap is recommended |
| | | |
Also available as EDU Edition (for educational institutions only): |
MCR 72 or MCR 92 plus accessories with a special academic discount |
Free EDU Toolbox including lab equipment and educational material |
| | | |