ru-RUuk-UAen-GB

MHC-trans

Full description
Compared to traditional sample stages for one sample only, the simultaneous conditioning of several samples significantly speeds up investigations on crystal structure changes induced by humidity and temperature.
The possibility for transmission XRD makes MHC-trans ideal for in-situ X-ray diffraction studies of pharmaceuticals and other organic materials.

Accurate control of sample conditions
Sample temperature control from -10 °C to 150 °C
Humidity control from 5 %rh to 95 %rh between 10 °C to 60 °C and up to 80 °C with maximum 70 %rh
Precise sample conditioning by indirect sample heating and uniform filling of the sample compartment with humid gas

Integrated sample changer for increased throughput
Simultaneous investigation of 8 samples
Highly comparable results with all samples analyzed under the same conditions
Easy loading of samples

High data quality with transmission X-ray geometry
Sample chamber designed for transmission XRD with focused beam for optimum data quality
Integrated primary beam stop
Little X-ray background from the sample stage components
Product specification
Model MHC-trans
Temperature range -10 °C to 150 °C
Humidity range 5 %rh to 95 %rh (70 %rh)
from 10 °C to 60 °C (80 °C)
Number of samples 8
Scan range 2 °2Θ to 55 °2Θ
Transmission geometry
Atmospheres Nitrogen, air, water vapor, inert gas
Pressure max. 1 bar rel.
Sample holder diameter 11 mm
Sample holder depth 1 mm / other on request
Carrier foil materials polysulfone or polyimide
Downloads
MHC-trans
MHC-trans
Платформи для рентгенівської дифракції.pdf

To get information about: «MHC-trans» contact us by phone (044) 229 15 31 or fill in the form.
E-mail
Phone
First name
Family name
Country
Company name
Request
Input symbols from the picture