A furnace heater provides uniform sample heating and excellent temperature control. Despite the extreme operating conditions, the sample can be exchanged easily without any tools. HPC 900 opens up completely new possibilities in hydrogen energy research and catalyst development.

Accurate control of sample conditions
Gas pressure up to 100 bar
Sample temperature from 25 °C to 900 °C
Can be operated with a variety of reactive gases, including hydrogen
Fast and uniform sample heating
Reliable temperature measurement right at the sample position

Safe and easy to use
Fast sample exchange without any tools
Removable sample cups for easy sample preparation
Gas-flushed safety shell around the pressurized sample compartment
UL and CE certified

Compact design
Robust and compact design using high quality materials
Fits on most available laboratory diffractometers
Available with automated z-alignment stage for temperature-dependent sample realignment
Product specification
Model HPC 900
Operating temperature 25 °C to 900 °C (depending on gas type and pressure)
Operating pressure 1 bar to 100 bar
Atmospheres H2, N2, O2,  inert gases
2θ range 0 – 160 ° 2θ
Chamber transmission 60 % for Mo Kα radiation
Sample holder Alumina (Al2O3) or Inconel 
Ø10 mm, depth 0.8 mm or 0.4 mm
Measure geometry Reflection