Every CO₂ incubator from BINDER guarantees ideal, reproducible growth conditions for cell cultures. The proven APT.line™ air jacket system and Permadry technology eliminate the risk of condensation. Thanks to 180 °C hot-air sterilization, bacteria and spores are eliminated.
Important characteristics
• Temperature range: room temperature plus 7 °C to 60 °C
• Humidity range: up to 95% RH
• Auto-sterilization with hot air at 180 °C
• Double-pan humidification system with condensation protection
• CO₂ gas-mixing jet with Venturi effect
• Hot-air sterilizable CO₂ sensor with infrared technology
• Humidity sensor with infrared technology
• Display via color LCD monitor
• Lockable door handle
• Tightly-sealed inner door made of tempered safety glass
• Seamless, deep-drawn inner stainless steel chamber
• 2/3 perforated stainless steel shelves
• Units are stackable with stacking adapter
• Troubleshooting system with visual and audible alarms
• Computer interface: Ethernet
• Internal data recording and USB interface
• Zero-voltage alarm contact
| | | |
Model | CB 60 | CB 160 | CB 220 |
Outer dimensions |
Width net [mm] | 580 | 680 | 740 |
Height net [mm] | 720 | 920 | 1070 |
Depth net [mm] | 545 | 715 | 715 |
Wall clearance back [mm] | 100 | 100 | 100 |
Wall clearance sidewise [mm] | 50 | 50 | 50 |
Inner doors | 1 | 1 | 1 |
Unit doors | 1 | 1 | 1 |
| | | |
Internal Dimensions |
Width [mm] | 400 | 500 | 560 |
Height [mm] | 400 | 600 | 750 |
Depth [mm] | 330 | 500 | 500 |
Interior volume [L] | 53 | 150 | 210 |
Net weight of the unit (empty) [kg] | 60 | 107 | 121 |
Load per rack [kg] | 10 | 10 | 10 |
| | | |
Performance Data Temperature |
Temperature range 7 °C above ambient temperature to [°C] | 60 | 60 | 60 |
Temperature uniformity at 37 °C [± K] | 0.3 | 0.3 | 0.4 |
Temperature fluctuation at 37 °C [± K] | 0.1 | 0.1 | 0.1 |
Recovery time after 30 seconds door open at 37 °C [min] | 4 | 4 | 5 |
Recovery time after 30 seconds door open at 5 Vol.-% CO₂ [min] | 5 | 5 | 5 |
| | | |
Performance Data Climate |
Humidity range [% RH] | 90…95 | 90…95 | 90…95 |
| | | |
Gas Data - CO₂-Data |
CO₂ range | 0…20 Vol.-% CO₂ | 0…20 Vol.-% CO₂ | 0…20 Vol.-% CO₂ |
CO₂ measurement | IR | IR | IR |
| | | |
Gas Data - O₂-Data |
O₂ range | 0,2…95 Vol-% O₂ | 0 Vol-% O₂ | 0 Vol-% O₂ |
| | | |
Electrical data |
Rated Voltage [V] | 200…240 | 200…240 | 200…240 |
Power frequency [Hz] | 50/60 | 50/60 | 50/60 |
Nominal power [kW] | 1 | 1.3 | 1.5 |
Unit fuse [A] | 10 | 10 | 10 |
Phase (Nominal voltage) | 1~ | 1~ | 1~ |
Energy consumption at 37 °C | 80 Wh/h | 100 Wh/h | 120 Wh/h |
| | | |
| | | |
CB-60
Технические характеристики шкафа CB-60
CB-160
Технические характеристики шкафа CB-160
CB-220
Технические характеристики шкафа CB-220
To get information about: «CO₂ incubators Series CB BINDER » contact us by phone (044) 229 15 31 or fill in the form.