Every CO₂ incubator from BINDER guarantees ideal, reproducible growth conditions for cell cultures. The proven APT.line™ air jacket system and Permadry technology eliminate the risk of condensation. Thanks to 180 °C hot-air sterilization, bacteria and spores are eliminated.

Important characteristics
• Temperature range: room temperature plus 7 °C to 60 °C
• Humidity range: up to 95% RH
• Auto-sterilization with hot air at 180 °C
• Double-pan humidification system with condensation protection
• CO₂ gas-mixing jet with Venturi effect
• Hot-air sterilizable CO₂ sensor with infrared technology
• Humidity sensor with infrared technology
• Display via color LCD monitor
• Lockable door handle
• Tightly-sealed inner door made of tempered safety glass
• Seamless, deep-drawn inner stainless steel chamber
• 2/3 perforated stainless steel shelves
• Units are stackable with stacking adapter
• Troubleshooting system with visual and audible alarms
• Computer interface: Ethernet
• Internal data recording and USB interface
• Zero-voltage alarm contact
Product specification
Model CB 60 CB 160 CB 220
Outer dimensions
Width net [mm] 580 680 740
Height net [mm] 720 920 1070
Depth net [mm] 545 715 715
Wall clearance back [mm] 100 100 100
Wall clearance sidewise [mm] 50 50 50
Inner doors 1 1 1
Unit doors 1 1 1
Internal Dimensions
Width [mm] 400 500 560
Height [mm] 400 600 750
Depth [mm] 330 500 500
Interior volume [L] 53 150 210
Net weight of the unit (empty) [kg] 60 107 121
Load per rack [kg] 10 10 10
Performance Data Temperature
Temperature range 7 °C above ambient temperature to [°C] 60 60 60
Temperature uniformity at 37 °C [± K] 0.3 0.3 0.4
Temperature fluctuation at 37 °C [± K] 0.1 0.1 0.1
Recovery time after 30 seconds door open at 37 °C [min] 4 4 5
Recovery time after 30 seconds door open at 5 Vol.-% CO₂ [min] 5 5 5
Performance Data Climate
Humidity range [% RH] 90…95 90…95 90…95
Gas Data - CO₂-Data
CO₂ range 0…20 Vol.-% CO₂ 0…20 Vol.-% CO₂ 0…20 Vol.-% CO₂
CO₂ measurement IR IR IR
Gas Data - O₂-Data
O₂ range 0,2…95 Vol-% O₂ 0 Vol-% O₂ 0 Vol-% O₂
Electrical data
Rated Voltage [V] 200…240 200…240 200…240
Power frequency [Hz] 50/60 50/60 50/60
Nominal power [kW] 1 1.3 1.5
Unit fuse [A] 10 10 10
Phase (Nominal voltage) 1~ 1~ 1~
Energy consumption at 37 °C 80 Wh/h 100 Wh/h 120 Wh/h
Downloads
CB-60
Технические характеристики шкафа CB-60
CB-160
Технические характеристики шкафа CB-160
CB-220
Технические характеристики шкафа CB-220
Multimedia
СО2 - інкубатори BINDER
СО2 - інкубатори BINDER


СО2 - інкубатори BINDER
СО2 - інкубатори BINDER



To get information about: «CO₂ incubators Series CB BINDER » contact us by phone (044) 229 15 31 or fill in the form.
E-mail
Phone
First name
Family name
Country
Company name
Request
Input symbols from the picture