Model | B 28 |
Measures |
Interior volume [L] | 28 |
Net weight of the unit (empty) [kg] | 23 |
Permitted load [kg] | 25 |
Load per rack [kg] | 10 |
Internal Dimensions |
Width [mm] | 400 |
Height [mm] | 280 |
Depth [mm] | 250 |
Doors |
Unit doors | 1 |
Outer dimensions |
Width net [mm] | 580 |
Height net [mm] | 405 |
Depth net [mm] | 425 |
Wall clearance back [mm] | 160 |
Wall clearance sidewise [mm] | 100 |
Fixtures |
Number of shelves (std./max.) | 2/4 |
Performance Data Temperature |
Temperature range [°C] | 30..70 |
Temperature variation at 37 °C [± K] | 1.5 |
Temperature fluctuation at 37 °C [± K] | 1 |
Heating-up time to 37 °C [min] | 95 |
Electrical data |
Rated Voltage [V] | 230 |
Power frequency [Hz] | 50/60 |
Nominal power [kW] | 0.25 |
Phase (Nominal voltage) [ph] | 1~ |
| |
* model B 28 can be proposed with TB class 1 | |
All technical data is specified for unloaded units with standard equipment at an ambient temperature of +22 °C ±3 °C and a power supply voltage fluctuation of ±10%. The temperature data is determined in accordance to BINDER factory standard following DIN 12880, observing the recommended wall clearances of 10 % of the height, width, and depth of the inner chamber. All indications are average values, typical for units produced in series. We reserve the right to change technical specifications at any time. |
| |