ru-RUuk-UAen-GB

L-Sonic 5100/L-Sonic 6100

Full description
L-Sonic 5100
• Concentration measurement, interface detection, and product identification
• For use in hazardous environments, i.e. chemical industry, and hygiene applications in the beverage and pharma industries
• Available in e.g. SST, Hastelloy BC1, rhodium plating, and more
• Easy integration, no maintenance, no consumables required

L-Sonic 6100
• Concentration and oil content ratio (OCR) measurements
• Suitable for the automotive industry, test stand centers for cooling systems, and in the chemical industry
• Comprehensive database for concentration formulas available
• Easy integration, no maintenance, no consumables required

L-Sonic tells a story: a 30-year story of sound velocity sensor development at Anton Paar, culminating in the sleek L-Sonic 5100/6100 series, which delivers pin-point accuracy and constant quality control. The L-Sonic 5100 has a fork-type senor, the L-Sonic 6100 a tube-type sensor. Use them for inline concentration measurements, interface detection, or product identification. And rest assured: the sensors continuously monitor the product quality of liquids during production.

L-Sonic: compact, stand-alone, supremely accurate
The L-Sonic series’ digital signal processing produces remarkable repeatability of 0.005 m/s (fork type). And a temperature-compensated crystal oscillator provides a measurement time basis of phenomenal exactitude.
• Measurements uninfluenced by changes in pressure, flow rate, and viscosity
• High tolerance to bubbles
• Applicable for interface detection or product identification (fork-type sensor)
• Ideal for OCR measurement (tube-type sensor)

The right instrument for demanding applications and hazardous environments
The L-Sonic sensors can be applied to a wide range of samples and are resistant to aggressive media because their wetted parts (L-Sonic 5100) comprise a variety of different materials. The sensor is suited to hazardous environments, including those where there is a risk of explosion, and to demanding hygiene applications in the beverage and pharmaceutical industries. L-Sonic’s modularity means easy and cost-efficient integration. The tube-type sensor L-Sonic 6100 can be directly installed in small pipes, so the entire flow passes through the sensor, and the fork-type sensor L-Sonic 5100 is ideal for insertion into large pipes or tanks. Anton Paar’s L-Sonic is versatile and resilient. Its solutions are comprehensive and customized across industry:
• Hazardous environments (e.g. chemical industry)
• Hygiene applications (e.g. beverages, pharma, automotive)
• Full flow, inline, bypass, or tank installation

Save time: quick installation and start-up; no maintenance needed
With these sensors, you’ll save a ton of time. Pre-configured and adjusted in the factory, they come with the right setup and integrated application formulas – installation, start-up, and commissioning are quick, no on-site adjustments, maintenance, or consumables are needed.

An Anton Paar measuring solution – tailored to your specific needs
Anton Paar draws from our comprehensive database and application know-how to give you the information you need to measure the concentration of your liquids precisely. In our in-house laboratory we measure your samples and develop the formula best suited to your needs. We identify the ideal measuring method (density, sound velocity, or refractive index) for your sample, as well as the perfect concentration and temperature range.
No matter which industry you’re in, we help you determine the concentration of liquids accurately and efficiently. L-Sonic is the right choice for the chemicals, beverages, and petroleum sectors, and more.
Product specification
Model L-Sonic 5100 L-Sonic 6100
Measuring range 800 m/s to 2500 m/s 200 m/s to 1000 m/s
Material of  Stainless steel 1.4404 (316L) Stainless Steel 1.4404 (316L)
the wetted parts HASTELLOY® HYBRID-BC1® alloy  
  Monel 400  
  Rhodium-coated  
  Fork length:   
  56 mm  
  125 mm  
  Customer-specific  
Accuracy in adjusted range:  
Reproducability  0.1 m/s
sound velocity    
Reproducability  0.02 °C
temperature    
Repeatability 0.005 m/s 0.01 m/s
   
Process temperature  -25°C to 125 °C
CIP/SIP temperature  145 °C for max. 30 min.
and duration    
Ambient temperature   -25 °C to 65 °C without HMI
(for non Ex versions)  -20 °C to 55 °C with HMI
Process pressure According to flange up to 70 bar @ 125 °C or
 absolute specification 100 bar @ 50 °C
Recommended flow rate >0 m/s to 6 m/s
Communication Pico 3000:
  • Analog
  • HART
  • Modbus RTU
  • PROFIBUS DP
  • PROFINET IO
   
  mPDS 5:
  • PROFIBUS DP
  • PROFINET IO
  • Devicenet
  • Ethernet/IP
  • Modbus TCP
  • I/O Board
   
Process connections VARIVENT® N Tube end:
  VARIVENT® G OD 12 mm
  DIN 11851  
  EN 1092-1  
  ANSI B16.5  
Dimensions:  
Non-Ex version Depending on model 150 mm x 145 mm x 175 mm
 (L x W x H)    
Ex version (L x W x H) Depending on model 160 mm x 160 mm x 190 mm
Explosion-proof versions according to ATEX / IECEx / FM / UL / CSA