ru-RUuk-UAen-GB

HT100 Automatic Tablet Testing System

Full description
HT 100 automatically measures tablet weight, thickness, width, diameter or length and hardness for up to 10 different products or batches ( e.g. tablets, dragees, oblongs, effervescent tablets, or similar drag delivery systems).
Precise measuring of up to 5 parameters
Equipped with patented AutoAlignment™ technology
AutoSeparator™ technology for vibration-free tablet separation
Measuring range of up to 500 or 1000 N
Fulfills USP <1217> on 3-point calibration & tablet orientation

10-chamber Feeder
The optional batch feeder has 10 chambers. Each chamber can hold 200 samples or more (dependent upon size and shape). Chambers may be filled with different batches of tablets or combined as one large single batch. The feeder cover serves two purposes, namely to load samples into the chamber and to protect the tablets against dust. The feeder moves clockwise and transports the samples into the AutoSeparator™.

AutoSeparator™
The AutoSeparator™ separates all shapes and sizes of tablets without any manual user adjustments or special tools. Separation is achieved through a device based on soft rotation (instead of vibration). The AutoSeparator™ is removable and can be easily cleaned. A photoelectric sensor is positioned on the outside of the separator to detect each individual sample as it is deposited from the separator to the balance boat.

4-station Transporter with AutoAlignment™ Technology
The 4-station transporter feeds the tablet clockwise from each test station. The AutoAlignment™ device (patented) on the hardness testing station instantaneously aligns the sample in the correct measurement position. It guarantees that all common forms are positioned correctly.

Hardness Measuring Station
Equipped with the AutoAlignment™ device (patented), to position automatically and rapidly all common forms at exactly 90 degrees to the test jaw before testing. An automated cleaning device that protects the hardness station from tablet residues is included. The measuring plate is ball bearing supported for an easy adjustment and to ensure that it remains parallel to the test platform. The measurement principle of tablet hardness can be changed from “Constant Speed” (where the test jaw is moved at a constant speed adjustable from 10–200 mm/min.) or to “Constant Force” (where the test jaw increases force linearly adjustable from 10–100 N/sec).

Documentation and Reporting
The q-doc prolab® Quality Documentation Software package for Windows XP / Windows 7 and 8, permitting the storage of methods and test results for weight, thickness, width, diameter or length and hardness. The software is fully 21 CFR Part 11 and GAMP 4 compliant. The software is also available with LIMS capabilities and can be networked. The reported test results can be customized, printed out, stored in the database, validated and/or sent to a network secure location.
Product specification
Модель HT100
Measurement principles Constant Speed,                                Constant Force
Range  10 – 200 mm/min 
  10 – 100 N/sec
Measurement Range  
Hardness (Accuracy) 10 – 500 (+/-1 N);                                   
  10 – 1000 (+/-2 N)
Diameter/Length  0.00 – 30.00 mm 
Accuracy  +/-0.05 mm
Thickness  0.00 – 18.00 mm 
Accuracy  +/-0.05 mm
Weight (Accuracy) 50 mg – 5 g (+/-0.2 mg);   
  5 g – 30 g (+/-1 mg )
Other
Power  230 V/50 Hz or 110V/60 Hz 
Width/depth/height  503/483/560 mm 
Weight  Approx. 39 kg 
Downloads
HT 100
Brochure in English: HT 100 Automatic Tablet Testing System

To get information about: «HT100 Automatic Tablet Testing System» contact us by phone (044) 229 15 31 or fill in the form.
E-mail
Phone
First name
Family name
Country
Company name
Request
Input symbols from the picture