• Built-in Peltier temperature control (20 °C or 25 °C)
• Small footprint; saves space in the lab
• Wireless automatic identification of sample cells and quartz control plates via RFID
The MCP 100 polarimeter is the right choice for routine quality control and universities. The easy-to-use touchscreen supports fast navigation with unparalleled simplicity. The powerful Peltier temperature control guarantees fast and even temperature distribution. With the intelligent Toolmaster™ technology, the connected sample cell or quartz control plate is automatically detected wirelessly. This eliminates user errors and ensures that every change is recorded in the audit trail. LIMS and FTP connectivity are fully supported.
Whenever lab bench space is limited, MCP 100 is the right choice
• Small footprint (less than two notebook pages)
• Lightweight instrument with enormous robustness
• Ideal for multiparameter setup, e.g. combination with a refractometer or density meter
Designed for ultimate ease of use
• Built-in Peltier temperature control
• Wireless identification of quartz plates and sample cells
• Temperature is wirelessly transferred from the sample cell to the instrument
• Color touchscreen for easy navigation
Fulfills regulations for quality management and international compliance
• Compliance to national and international pharmacopeias
• Outstanding feature set for your individual requirements
• Password protection for maximum safety
• User level management for easy access control
• Audit trail for full traceability
| |
Model | МСР 100 |
| |
Measuring scales: | °Optical Rotation, °Specific Rotation, %Concentration |
Measuring range: | ±89.9° |
Resolution: | 0.001° |
Accuracy: | ±0.01° |
Repeatability: | ±0.01° |
Response time: | 12 sec. |
Wavelength: | 589 nm |
Light source: | LED, with average lifetime of 100.000 hours |
Sensitivity: | A light intensity control compensates attenuation up to Optical Density (OD) 2.0 |
Temperature sensor: | Built-in PT 100 sensor |
Temperature resolution: | 0.1 °C |
Temperature accuracy: | ±0.2 °C |
Temperature control: | Automatic Peltier temperature control at 20°C and 25°C, or without temp. control |
Sample cell / quartz control plate: | ToolmasterTM - Automatic sample cell and quartz control plate identification; sample cell path length of 2.5 mm to 100 mm |
Display: | TFT touchscreen 5.7" (145 mm), 640 x 480 Pixels |
Interfaces: | 3 USB, RS232, Ethernet, CAN bus, easy connection of keyboard, mouse, printer, bar code. Network capability |
Dimensions (LxWxH): | 370 mm x 320 mm x 130 mm (14.6 in x 12.6 in x 5.1 in) |
Weight: | 7.8 kg (17.2 lbs) |
Power supply / consumption: | AC 100 to 240 VAC, 50/60Hz / 70VA, 120 VA max. |
| |
| |