A revolutionary measuring principle for the market-leading density and sound velocity meter
The unique design of the measuring cell, a new way of evaluating the oscillation characteristics, and many other innovative features make DSA 5000 M one of a kind
• Measures the whole concentration range of sulfuric acid and oleum
• Determines the concentration of two- and three-component solutions
• Also ideal for phosphoric acid, formaldehyde, methanol, inks, and paints
• Minimum sample volume increases safety for lab personnel
DSA 5000 M is the only instrument that combines density and sound velocity measurements in one setup. It determines the concentration of two- and three-component solutions using the most accurate density results in the market, measured with the Repeated Fade-out Method. R&D departments in many different industries already make use of the opportunity to measure both parameters in one go. What’s more: It is the only way to determine the concentration of sulfuric acid over the whole concentration range with just one instrument.
The best in accuracy, repeatability, and reproducibility
DSA 5000 M provides the best results in the market:
Density: Repeatability up to the 7th digit
Sound velocity: Repeatability up to 0.1 m/s
Reproducibility: 0.000005 g/cm³ (density), 0.5 m/s (sound velocity)
Eliminates temperature drifts
The ThermoBalance™ technology makes sure that this accuracy is reached under any temperature up to 100 °C. A reference oscillator is used to eliminate temperature-related fluctuations of the glass tube – even after years of operation. This elimination of drifts makes ThermoBalanceTM the ideal foundation for:
• Thermodynamical R&D studies with frequently changing measuring temperatures
• Routine quality control with changing sample temperatures
The fastest way to determine sulfuric acid concentration
In terms of accuracy, DSA 5000 M is the best solution for sulfuric acid. The combination of density and sound velocity measurements makes sure that you notice the smallest deviations in your production process. You can react quickly, need less time to correct errors, and produce less rejects. DSA 5000 M is also the fastest way to measure sulfuric acids. It is the only way to measure sulfuric acid over the whole concentration range in one instrument – with no need to change setups in between different samples or to use tremendous amounts of consumables such as reagents. Whether the concentration of your sulfuric acid is 0 % or 114 %, DSA 5000 M delivers the result in less than two minutes.
Detects filling errors and gas bubbles
FillingCheckTM detects filling errors and generates a warning if it finds gas bubbles or particles in your sample. If you filled your sample correctly, the feature will give you a “go” to start the measurement. Additionally, you can always check the U-ViewTM camera that displays and stores live images of the filled U-tube. This real-time monitoring notifies you of mistakes right away and gives you the peace of mind that your results are not distorted by filling errors.
One sample, one measurement cycle, numerous results
DSA 5000 M retrieves several parameters from one and the same sample.
Measured parameters:
• Density
• Sound velocity
Derived measurands:
• Concentration of two-component solutions
• Concentration of three-component solutions
• Adiabatic compressibility
The sample automatically runs through both measuring cells. For one run, you need only 3.5 mL of sample to get all these parameters.
The most stable results available
The viscosity correction is responsible for the high repeatability of density results in DSA 5000 M. It eliminates viscosity-related errors twice as effectively as ever before and works automatically across the entire viscosity range of your samples. Thanks to this technology, you can detect the smallest deviations in your continuous production and react quickly to minimize downtimes. The viscosity correction already comes calibrated ex-factory so you will never need viscosity standards, saving you time for calibration and money for standards.
Model | DSA 5000 M |
Patents |
Patent granted | AT 516420 (B1), AT 517082 (B1) |
Patents pending | AT 517486 (A1) |
Measuring range |
Density | 0 g/cm³ to 3 g/cm³ |
Sound velocity | 1000 to 2000 m/s |
Temperature range: | 0 °C to 100 °C |
Pressure range | 0 to 8 bar |
Accuracy |
Density: | 0.000007 g/cm³ |
Concentration determination | Typically 0.01 to 0.1 % (application-dependent) |
Repeatability** |
Density | 0.000001 g/cm3 |
Sound velocity | 0.1 m/s |
Temperature | 0.001 °C |
0 to 100 % H2SO4 | 0.02 % H2SO4 |
0 to 28 % free SO3 | 0.04 % free SO3 |
28 to 65 % free SO3 | 0.1 % free SO3 |
Digital resolution |
Density | 0.000001 g/cm3 |
Sound velocity | 0.01 m/s |
Additional information |
Integrated tables and functions | Ethanol tables |
| Extract/sugar tables |
| Acid/base tables |
| 20 freely programmable user functions |
Available options | ISO 17025 calibration providing full traceability to SI units |
| Cooling kit for low-temperature measurement |
Amount of sample in the | 3 mL |
measuring cells | |
Measuring time per sample | 1 to 4 minutes |
Interfaces | RS-232, 4 x USB, CAN, VGA, Ethernet |
Display | Bright 10.4" TFT PCAP touchscreen (640 x 480 px) with |
| customizable display layout |
Wetted parts | PTFE, borosilicate glass, stainless steel DIN 1.4539/UNS |
| N08904 |
Dimensions | 495 x 330 x 230 mm |
Weight | 22.5 kg (49.6 lbs) |
*under ideal conditions and for low densities/viscosities |
**according to ISO 5725 | |
| |