| Model | Xsample 530 |
| Viscosity range | |
| • Operation with internal air pump | 12.000 mPa·s at ambient and measuring temperature |
| • Operation with 2 bar external compressed air | 36.000 mPa·s at ambient and measuring temperature |
| Filling mode | Pressure mode |
| | • with integrated piston pump |
| | • with air pump of master instrument |
| | • with compressed air |
| | Suction mode (optional) |
| Minimum sample volume per measurement | approx. 3 mL (with master instrument DMA™ M) |
| | depending on system configuration |
| Sample recovery | optional |
| Rinse consumption per cycle | approx. 4 mL |
| Sample throughput | max. 9 per hour |
| | depending on application and used pressure source |
| Number of sample vials | 71 positions for 12 ml vials |
| | optional: 35 positions for 45 ml vials |
| | customer specific magazines on request |
| Automatic filling and drying | Piston pump, internal air pump of master instrument or compressed air selectable |
| Compressed air supply | 1.80 bar – 2.20 bar (26 psi – 32 psi) relative external compressed air |
| (optional) | typically: 2.0 bar (29 psi) rel. |
| Necessary quality of air | Class 3 from ISO 8573-1: |
| | max. particle size: 5 μm |
| | max. pressure dew point:-20°C (-4°F) |
| | max. oil content: 1 mg/m3 |
| Environmental conditions (EN 61010) | Indoor use only |
| Air humidity | 10 to 90% relative humidity, non-condensing |
| Operating temperature | +15 to +35 °C (+59 to +95 °F) |
| Power supply | supplied by master instrument |
| | Mains voltage according to the type plate at the rear of master instrument (protective grounding required) |
| Dimensions excl. master instrument | 370 mm x 490 mm x 430 mm |
| (L x W x H) | (14.6 in x 19.3 in x 17 in) |
| Weight excl. master instrument | approx. 48 kg (105.8 lbs) |
| Optional bar code reader | Laser Line Scanner, 1D Code Reader class 2 |
| |