Important characteristics
Temperature range: room temperature plus 5 °C to 300 °C
APT.line™ preheating chamber technology
Controller with time-segment and real-time programming
2 chrome-plated racks
Fresh-air monitoring with audible and visual alarm and automatic deactivation of heating
Replaceable filter cartridge, Class F6
Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm
Computer interface: RS 422
Model | FED 53 |
Outer dimensions | |
Width net [mm] | 830 |
Height net [mm] | 805 |
Depth net [mm] | 685 |
Wall clearance sidewise [mm] | 100 |
Wall clearance back [mm] | 160 |
| |
Internal Dimensions | |
Width [mm] | 600 |
Height [mm] | 435 |
Depth [mm] | 435 |
Interior volume [L] | 115 |
Shelves (standard/max) | 2 / 5 |
Load per rack [kg] | 30 |
Full load [kg] | 50 |
Net weight of the unit (empty) [kg] | 90 |
| |
Performance Data Temperature | |
Temperature range 10 °C above ambient temperature to [°C] | 300 |
Temperature uniformity at 150 °C [± K] | 2.5 |
Temperature fluctuation [± K] | 0.8 |
Heating-up time to 150 °C [min] | 18 |
Recovery time after 30 seconds door open at 150 °C [min] | 7 |
Air circulation (approx.) [x/min] | 20 |
Volumetric flow rate of exhaust air acc. to EN 1539 at 50 °C [ca. L/min] | 400 |
Vakuumdaten | |
Highest permitted solvent quantity (at T-180 °C, M-100g/mol, U-40g/m3, K=0,5) [g] | 6.65 |
| |
Electrical data | |
Rated Voltage [V] | 230 |
Power frequency [Hz] | 50/60 |
Nominal power [kW] | 2.9 |
Phase (Nominal voltage) | 1~ |
| |
Environment-specific data | |
Energy consumption at 150 °C | 1200 Wh/h |
Sound-pressure level [dB(A)] | 57 |
| |
FDL-115
Технические характеристики шкафа FDL-115
Брошюра сушильные шкафы и нагревательные камеры