ru-RUuk-UAen-GB

Avantgarde.Line Series ED with natural convection

Full description

  • Temperature range: ambient temperature plus 5 °C to 300 °C

  • Up to 30% lower energy consumption compared to conventional units on the market

  • High temperature accuracy thanks to APT.line™ technology

  • Natural convection

  • Controller with LCD display

  • Electromechanical control of the exhaust air flap

  • 2 chrome-plated racks

  • Units up to 115 liters are stackable

  • Class 2 integrated independent adjustable temperature safety device (DIN 12880) with visual alarm

  • Ergonomic handle design

  • USB port for recording data

Product specification
Model ED 56 ED 115 ED 260 ED 720
Measures
Interior volume [L] 57 114 255 743
Net weight of the unit (empty) [kg] 39 54 85 169
Permitted load [kg] 70 150 270 315
Load per rack [kg] 30 30 40 45
Internal Dimensions
Width [mm] 360 510 610 960
Height [mm] 420 530 760 1280
Depth [mm] 380 425 550 605
Doors
Unit doors 1 1 1 2
Inner doors - - - 2
Outer dimensions
Width net [mm] 560 710 810 1165
Height net [mm] 625 735 965 1590
Depth net [mm] 565 605 760 870
Wall clearance back [mm] 160 160 160 160
Wall clearance sidewise [mm] 100 100 100 100
Environment-specific data
Energy consumption at 150 °C [Wh/h] 180 245 355 700
Fixtures
Number of shelves (std./max.) 2/4 2/5 2/8 2/16
Performance Data Temperature
Temperature range 5 °C above ambient temperature to [°C] 300 300 300 300
Temperature variation at 150 °C [± K] 2.3 1.5 1.8 3.2
Temperature fluctuation at 150 °C [± K] 0.4 0.4 0.8 0.8
Heating-up time to 150 °C [min] 50 45 50 85
Recovery time after 30 seconds door open at 150 °C [min] 19 16 20 25
Electrical data
Rated Voltage [V] 230 230 230 400
Power frequency [Hz] 50/60 50/60 50/60 50/60
Nominal power [kW] 1 1 2.25 4
Unit fuse [A] 6.3 6.3 12.5 16
Phase (Nominal voltage) [ph] 1~ 1~ 1~ 3~
All technical data is specified for unloaded units with standard equipment at an ambient temperature of +22 °C ±3 °C and a power supply voltage fluctuation of ±10%. The temperature data is determined in accordance to BINDER factory standard following DIN 12880, observing the recommended wall clearances of 10 % of the height, width, and depth of the inner chamber. All indications are average values, typical for units produced in series. We reserve the right to change technical specifications at any time.
Downloads
Брошюра сушильные шкафы и нагревательные камеры
Брошюра сушильные шкафы и нагревательные камеры
Avantgarde.Line ED 115
Binder Avantgarde.Line ED 115
Avantgarde.Line ED 720
Binder Avantgarde.Line ED 720

To get information about: «Avantgarde.Line Series ED with natural convection» contact us by phone (044) 229 15 31 or fill in the form.
E-mail
Phone
First name
Family name
Country
Company name
Request
Input symbols from the picture